
Support
Frequently Asked Questions of Firstrate Pressure Sensor
- Categories:Technical support
- Time of issue:2016-10-28 19:16
(Summary description)1. After pressurizing, the output of the transmitter does not change? Answer: In this case, first check whether the pressure port is leaking or blocked. If it is not confirmed, check the wiring method. If the wiring is correct, check the power supply. If the power supply is normal, check whether the sensor zero has output, or perform a simple addition. Look at whether the output changes. If there is a change, the sensor is not damaged. If there is no change, the sensor has been damaged. Other reasons for this situation may also be damage to the instrument or other problems in the entire system. 2. The output of the pressure transmitter does not change, and the output of the pressure transmitter changes suddenly, and the zero position of the pressure relief transmitter cannot return? Answer: The reason for this phenomenon is most likely caused by the pressure sensor sealing ring. Generally, it is due to the specifications of the sealing ring (too soft or too thick). When the sensor is tightened, the sealing ring is compressed into the pressure port of the sensor to block the sensor. The pressure medium cannot enter when the pressure is high, but when the pressure is very high, the sealing ring is suddenly opened, and the pressure sensor is changed by the pressure. When the pressure drops again, the sealing ring returns to block the pressure port, and the remaining pressure cannot be released. Therefore, the zero position of the sensor cannot be lowered. The best way to eliminate this reason is to remove the sensor and directly check whether the zero position is normal. If it is normal, replace the sealing ring and try again. 3. The output signal of the transmitter is unstable? Answer: There are several reasons for the unstable output signal: A. The pressure source itself is an unstable pressure; B. The anti-interference ability of the instrument or pressure sensor is not strong (check whether there are interference sources around, such as motors, inverters, etc.); C. The sensor wiring is not firm (the cable wiring is bad or damaged, and the wiring needs to be re-wired); D. Check the grounding condition of the equipment (the induced voltage will charge the sensor or the instrument casing); E. Sensor failure; 4. There is no output when the transmitter is connected to electricity? Answer: The possible reasons are: A. Wrong wiring (check both the instrument and the sensor); B. Open circuit or short circuit of the wire itself; C. The power supply has no output or the power supply does not match (check whether the input power supply meets the power supply range); D. The meter is damaged or the meter does not match; E. The sensor is damaged; 5. After the sensor is connected to the meter, there is no output display? Answer: The sensor is connected to the instrument, and there is no output display. The sensor and the instrument can be tested separately to confirm which part of the fault is; Disconnect the signal output line of the sensor and the instrument, first check whether the sensor output is normal; The current signal and voltage signal output can be detected separately according to the above figure. If the zero position is normal, slowly pressurize the sensor to see if the output value changes. If the output value changes, it means the sensor is normal and it can be judged as a failure of the instrument. (This method can also be used for sensors with mV signal output). The sensor can be powered by a separate external power supply or a meter to see if the sensor output is normal. 6. How does the installation position of the differential pressure transmitter affect the zero output? Answer: Due to the small measuring range of the differential pressure transmitter, the weight of the sensing element in the transmitter will affect the output of the differential pressure transmitter. Therefore, the zero position of the differential pressure transmitter will appear when the differential pressure transmitter is installed. Changes are normal. During installation, the pressure sensitive part of the transmitter should be made perpendicular to the direction of gravity. If the installation conditions are limited, the zero position of the transmitter should be adjusted to the standard value after installation and fixation. 7. Transmitter output ≥20mA? Answer: The possible reasons are: A. Check whether the actual pressure exceeds the selected range of the pressure transmitter; B. Caused by overload of the sensor (severe overload sometimes damages the isolation diaphragm); C. Sensor failure; 8. The FST500-202 electronic pressure switch is not pressurized and the display is not zero? Answer: The possible reason is that the display of non-pressurization is not zero. A. After the product is installed, there may be slight pressure on the equipmen
Frequently Asked Questions of Firstrate Pressure Sensor
(Summary description)1. After pressurizing, the output of the transmitter does not change?
Answer: In this case, first check whether the pressure port is leaking or blocked. If it is not confirmed, check the wiring method. If the wiring is correct, check the power supply. If the power supply is normal, check whether the sensor zero has output, or perform a simple addition. Look at whether the output changes. If there is a change, the sensor is not damaged. If there is no change, the sensor has been damaged.
Other reasons for this situation may also be damage to the instrument or other problems in the entire system.
2. The output of the pressure transmitter does not change, and the output of the pressure transmitter changes suddenly, and the zero position of the pressure relief transmitter cannot return?
Answer: The reason for this phenomenon is most likely caused by the pressure sensor sealing ring. Generally, it is due to the specifications of the sealing ring (too soft or too thick). When the sensor is tightened, the sealing ring is compressed into the pressure port of the sensor to block the sensor. The pressure medium cannot enter when the pressure is high, but when the pressure is very high, the sealing ring is suddenly opened, and the pressure sensor is changed by the pressure. When the pressure drops again, the sealing ring returns to block the pressure port, and the remaining pressure cannot be released. Therefore, the zero position of the sensor cannot be lowered. The best way to eliminate this reason is to remove the sensor and directly check whether the zero position is normal. If it is normal, replace the sealing ring and try again.
3. The output signal of the transmitter is unstable?
Answer: There are several reasons for the unstable output signal:
A. The pressure source itself is an unstable pressure;
B. The anti-interference ability of the instrument or pressure sensor is not strong (check whether there are interference sources around, such as motors, inverters, etc.);
C. The sensor wiring is not firm (the cable wiring is bad or damaged, and the wiring needs to be re-wired);
D. Check the grounding condition of the equipment (the induced voltage will charge the sensor or the instrument casing);
E. Sensor failure;
4. There is no output when the transmitter is connected to electricity?
Answer: The possible reasons are:
A. Wrong wiring (check both the instrument and the sensor);
B. Open circuit or short circuit of the wire itself;
C. The power supply has no output or the power supply does not match (check whether the input power supply meets the power supply range);
D. The meter is damaged or the meter does not match;
E. The sensor is damaged;
5. After the sensor is connected to the meter, there is no output display?
Answer: The sensor is connected to the instrument, and there is no output display. The sensor and the instrument can be tested separately to confirm which part of the fault is;
Disconnect the signal output line of the sensor and the instrument, first check whether the sensor output is normal;
The current signal and voltage signal output can be detected separately according to the above figure. If the zero position is normal, slowly pressurize the sensor to see if the output value changes. If the output value changes, it means the sensor is normal and it can be judged as a failure of the instrument. (This method can also be used for sensors with mV signal output). The sensor can be powered by a separate external power supply or a meter to see if the sensor output is normal.
6. How does the installation position of the differential pressure transmitter affect the zero output?
Answer: Due to the small measuring range of the differential pressure transmitter, the weight of the sensing element in the transmitter will affect the output of the differential pressure transmitter. Therefore, the zero position of the differential pressure transmitter will appear when the differential pressure transmitter is installed. Changes are normal. During installation, the pressure sensitive part of the transmitter should be made perpendicular to the direction of gravity. If the installation conditions are limited, the zero position of the transmitter should be adjusted to the standard value after installation and fixation.
7. Transmitter output ≥20mA?
Answer: The possible reasons are:
A. Check whether the actual pressure exceeds the selected range of the pressure transmitter;
B. Caused by overload of the sensor (severe overload sometimes damages the isolation diaphragm);
C. Sensor failure;
8. The FST500-202 electronic pressure switch is not pressurized and the display is not zero?
Answer: The possible reason is that the display of non-pressurization is not zero.
A. After the product is installed, there may be slight pressure on the equipmen
- Categories:Technical support
- Time of issue:2016-10-28 19:16

v
Contact us
Tel:
+86 731-86171990
Email:
sales@firstratesensor.com
Address: 4th Floor, Building 1, Zhitingyuan, Zhenhua Road, Yuhua District, Changsha City, Hunan Province,China.
